Serie VA BUS

Product no.: GMF VA BUS (Gas type: further on request, Measuring range: further on request)

BUS-sensor for wet areas in stainless steel housing (chloride-resistant)

Measuring prinziple: electrochemical
Messgenauigkeit: <+- 1% of measuring range
Response time T90: < 30 sec
Ambient: -20..+50°C, 0..100% rF, 700 - 1300 hPa
Suitability: food industry, meat-processing industry,
wet rooms with pressure washer use,
breweries, wineries, washing tunnels
sewage plant
Housing: stainless steel, rust and acid resistant
LxWxD: 110x110x70 mm
: housing: IP66 K
Gas entry: diffusion, teflon filter
Output signal: digital, RS485
CE-conformidity: emission: residential: immunity for industrial environments
Supply: 12..36V DC / 4 W
Connection cable: up to 1200 m: JY(St)Y 2x2x0,8 mm
   
 

Operating instruction

Selection: Serie VA BUS

Product no. Gas type Measuring range Price
GMF 435 E CLO2 1 VA Chlorine dioxide 0...1000 ppb
GMF 435 E H2S 20 VA Hydrogen sulfide 0...20 ppm
GMF 735 IR CO2 10V VA Carbon dioxide 0...10 vol%
GMF 735 IR CO2 5V VA Carbon dioxide 0...5 vol%
GMF 735 IR HC UEG VA Hydrocarbon 0...100% LEL
GMF 735 IR PRP UEG VA Propane 0...100% LEL
GMF VA BUS further on request further on request